An audit of marker placement in stereotactic guided biopsy

Introduction

Anecdotal evidence suggests that there is a greater incidence of marker migration using large volume sampling techniques in stereotactic guided breast biopsies.

Methods

Prospective study of 130 biopsies with markers done between June and December 2014. Markers more than 10 mm from the target lesion were considered migrated. The aim of the audit was to quantify the number of markers migrating, distance and direction of migration and conditions under which markers migrate.

Results

A total of 12.3 % had migrated markers: 10.7 % from use of the Bard Encor system and 1.5 % from use of the Bard Vacora system. The greatest marker migration occurred using a latero-medial approach. The majority of migrated markers were deeper than the target lesion. Marker migration was significantly greater using the Encor system within lucent breast tissue. Firstly, further audit is required incorporating lesion size, routine vacuuming of the cavity before deployment of marker, specific sequencing of marker films, correlation of compressed breast thickness and target depth, clinical impact of marker migration and possible development of expanding marker. Secondly, the breast screening service should provide guidelines regarding distances, thresholds and targets for marker migration.

Conclusion

This audit found that marker migration occurred predominantly within lucent breast tissue and using the latero-medial approach when using the Bard Encor system.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Prady, J., Hill, L., Gray, A. et al. An audit of marker placement in stereotactic guided biopsy. Breast Cancer Res 17 (Suppl 1), P26 (2015). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/bcr3788

Download citation

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/bcr3788

Keywords