Digital breast tomosynthesis improves the accuracy of the diagnosis of circumscribed lesions because of increase of margin visibility
Breast Cancer Research volume 16, Article number: O6 (2014)
Introduction
Circumscribed masses are a common cause of recall for assessment but most are benign. The aim of this study is to compare the accuracy of digital breast tomosynthesis (DBT) with digital mammography (DM) in assessing the margin of circumscribed lesions as a predictor of benign or malignant disease.
Methods
The study group consisted of women recalled from breast screening for further assessment of circumscribed masses. Clients underwent co-registered DM and DBT in both MLO and CC projections. Two experienced breast radiologists evaluated DM and DBT images and a consensus decision was reached on the percentage of the margin that was well defined on DM and DBT. The lesions were categorised 1 to 4 as follows: 1 = 0 to 25%, 2 = 26 to 50%, 3 = 51 to 75% and 4 = 76 to 100%.
Results
One hundred and twenty circumscribed lesions were evaluated. Data on 118 lesions seen on the MLO view are presented. There were 93 benign lesions and 25 cancers. There was a change in distribution of margin categories between DM and DBT. More lesions were categorised as 3 or 4 on DBT (59/118) compared with DM (18/118). Of the 93 benign lesions, 17 were categorised as 3 or 4 on DM and 57 on DBT. The difference between the two proportions was significant (P < 0.0002). There were more cancers categorised as 1 on DBT (21/39 = 54%) compared with DM (20/76 = 26%); Fisher's exact test (P < 0.004).
Conclusion
Increased margin visibility of circumscribed masses by DBT improves the accuracy of mammography interpretation and may decrease the recall rate in mammography screening.
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.
The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.
To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Wasan, R., Morel, J., Iqbal, A. et al. Digital breast tomosynthesis improves the accuracy of the diagnosis of circumscribed lesions because of increase of margin visibility. Breast Cancer Res 16 (Suppl 1), O6 (2014). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/bcr3701
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/bcr3701